How to use the 3M economic BD test package:

1. The sterilizer is preheated to a suitable temperature

2. In the case of an empty pot, place the test bag above the exhaust port and near the door. Label side up, flat

3. Run the standard BD test cycle according to national standards.

4. After the end of the cycle, check the BD test paper.

3M economic BD test package interpretation method:

Pass (negative): BD test paper becomes uniform dark brown to black for passing test results, at this time BD test passed

Failure (positive): There is a bright area in the center of the BD test paper, and the color unevenness is the test result of the failure. At this time, the BD test is unqualified and should be processed according to the requirements of the national standard.